Background: Navigational software provides real-time fluoroscopic needle guidance for percutaneous procedures in the Interventional Radiology (IR) suite.Objective: We describe our experience with navigational software for pediatric percutaneous bone biopsies in the IR suite and compare technical success, diagnostic accuracy, radiation dose and procedure time with that of CT-guided biopsies.Materials and Methods: Pediatric bone biopsies performed using navigational software (Syngo iGuide, Siemens Healthcare) from 2011 to 2016 were prospectively included and anatomically matched CT-guided bone biopsies from 2008 to 2016 were retrospectively reviewed with institutional review board approval. C-arm CT protocols used for navigational software-assisted cases included institution-developed low-dose (0.1/0.17 μGy/projection), regular-dose (0.36 μGy/projection), or a combination of low-dose/regular-dose protocols. Estimated effective radiation dose and procedure times were compared between software-assisted and CT-guided biopsies.Results: Twenty-six patients (15 male; mean age: 10 years) underwent software-assisted biopsies (15 pelvic, 7 lumbar and 4 lower extremity) and 33 patients (13 male; mean age: 9 years) underwent CT-guided biopsies (22 pelvic, 7 lumbar and 4 lower extremity). Both modality biopsies resulted in a 100% technical success rate. Twenty-five of 26 (96%) software-assisted and 29/33 (88%) CT-guided biopsies were diagnostic. Overall, the effective radiation dose was significantly lower in software-assisted than CT-guided cases (3.0±3.4 vs. 6.6±7.7 mSv, P=0.02). The effective dose difference was most dramatic in software-assisted cases using low-dose C-arm CT (1.2±1.8 vs. 6.6±7.7 mSv, P=0.001) or combined low-dose/regular-dose C-arm CT (1.9±2.4 vs. 6.6±7.7 mSv, P=0.04), whereas effective dose was comparable in software-assisted cases using regular-dose C-arm CT (6.0±3.5 vs. 6.6±7.7 mSv, P=0.7). Mean procedure time was significantly lower for software-assisted cases (91±54 vs. 141±68 min, P=0.005).Conclusion: In our experience, navigational software technology in the IR suite is a promising alternative to CT guidance for pediatric bone biopsies providing comparable technical success and diagnostic accuracy with lower radiation dose and procedure time, in addition to providing real-time fluoroscopic needle guidance.