Computational Evaluation of Suspended Microcantilever and Microfluidic Channel
- Resource Type
- Conference
- Authors
- Gavalas, Iakovos; Fotiadis, Dimitrios I.
- Source
- 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) Engineering in Medicine and Biology Society (EMBC), 2019 41st Annual International Conference of the IEEE. :1171-1174 Jul, 2019
- Subject
- Bioengineering
Biosensors
Viscosity
Microfluidics
Biological system modeling
Finite element analysis
Sensitivity
- Language
- ISSN
- 1558-4615
In this study, a static mode suspended microcantilever model was simulated employing the Finite Elements Method (FEM), to examine the deflection of the microcantilever’s beam using Newtonian and Non-Newtonian fluids. The scope of this work is to design and simulate a suspended microcantilever biosensor for the determination of the fluids’ dynamic viscosity and the relationship between the blood’s hematocrit and the biosensor’s sensitivity. These biosensors are microstructures which consist of a microfluidic channel embedded in a microcantilever biosensor. Fluid Structure Interaction (FSI) simulations performed to study the deflection of the microcantilever when Newtonian and Non-Newtonian fluids with different dynamic viscosity are used. For the dynamic viscosity determination of the Non-Newtonian fluids, the Brid-Carreau model was used. The results indicate a proportional relation between the beam’s deflection and the fluids’ dynamic viscosity.