Objectives To evaluate bare-metal Optimus and polytetrafluoroethylene (PTFE)-covered Optimus-CVS balloon-expandable, cobalt–chromium, hybrid cell–designed stents in congenital heart disease (CHD) interventions.Methods Retrospective multicentre review of patients with CHD receiving Optimus stents. Stent mechanical behaviour, clinical indications and outcomes were assessed.Results 183 stents (49.2% XXL/15-ZIG, 33.3% XL/12-ZIG, 17.5% L/9-ZIG) were implanted (98.9% success rate, 2.3% serious procedural complication rate) in 170 patients (57.6% men, 64.1% adults), median age 23.6 years (IQR, 15.2–39.2) and weight 63.5 kg (IQR, 47–75.7). Indications were right ventricular outflow tract stand-alone stenting or before revalvulation (62.4%), aortic coarctation treatment (15.3%), Fontan-circuit fenestration closure (12.4%) and miscellaneous (10%). 86/170 (50.6%) patients had PTFE-covered stenting (50% prophylactic). In 86/170 (50.6%) patients with stenotic lesions, median percentage of achieved stent expansion was 93.4% (IQR, 85.5%–97.7%), median gradient decreased from 28 mm Hg (IQR, 19–41) to 5 mm Hg (IQR, 1–9) (p