Introduction: The anterior maxillary infiltration is one of the more painful dental injections. The Dentapen is an electronic syringe that uses computer-controlled delivery technology to administer dental local anesthesia at a slow controlled rate. The purpose of this prospective, randomized, single-blind study was to evaluate solution deposition pain of a maxillary lateral incisor infiltration using the Dentapen with the slow flow rate (1.8 mL/162 sec) and ramp-up setting compared with a traditional syringe infiltration at a flow rate of 1.8 mL/60 sec.
Methods: One hundred thirty adults were administered a maxillary lateral incisor infiltration with the Dentapen and a traditional syringe at 2 separate appointments in a single-blind manner. The infiltrations of 2% lidocaine with 1:100,000 epinephrine were given at a rate of 1.8 mL/162 sec with the ramp-up feature for the Dentapen and 1.8 mL/60 sec for the traditional infiltration. The pain of solution deposition was recorded on a visual analogue scale. At the conclusion of the study, subjects selected their preferred injection technique. The data were analyzed statistically using paired t tests, a mixed-effect model, and odds ratio.
Results: The pain of solution deposition was significantly less for the Dentapen injection than the traditional injection (P < .001). With the Dentapen device, 16% experienced moderate pain, and for the traditional syringe, 39% experienced moderate pain. Overall, 75% of subjects preferred the Dentapen injection over the traditional injection.
Conclusions: The Dentapen, using the slow flow rate and ramp-up mode, significantly reduced the pain of solution deposition for maxillary lateral incisor infiltrations.
(Copyright © 2022 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.)