The need for a routinely applicable assay to measure low estradiol levels in adult men, postmenopausal women, and young adolescents was recently discussed in an Endocrine Society position statement. Our aim was to develop a sensitive liquid chromatography-tandem mass spectrometry method for estradiol and estrone in human serum without the need for derivatization or extended extraction protocols. After protein precipitation of serum with a mixture of methanol/acetonitrile (85/15) ( v/ v) containing isotopic internal standards (17β-estradiol-16,16,17- d and estrone-2,3,4-C), we quantified estradiol and estrone by two-dimensional liquid chromatography-tandem mass spectrometry with electrospray ionization in the negative mode monitoring 5 × 271.20→145.00 (17β-estradiol) and 269.20→145.00 (estrone). Sensitivity was increased by using fluoride and summation of 5 identical transitions for estradiol. Our method was analytically validated, compared against direct immunoassays using serum of 25 adult men, and clinically tested by measuring samples of 3 men at baseline and after chemical castration, 30 postmenopausal women and 15 patients receiving aromatase inhibitors. Total imprecision was below 20 % for the low quality controls. Limit of quantification was 1.3 ng/L (4.8 pmol/L) for estradiol and 1.2 ng/L (4.4 pmol/L) for estrone. Estradiol in Certified Reference Material BCR-576 was within specified uncertainty limits. No significant ion suppression or interference was observed. Our method showed modest correlation with direct immunoassay for estradiol ( r = 0.64) but no correlation for estrone ( r = 0.12). Patient sample results were within expected ranges. In conclusion, we developed a routinely applicable liquid chromatography-tandem mass spectrometry method for estradiol and estrone measurement which is sensitive enough for use in men, postmenopausal women, and young adolescents. [Figure not available: see fulltext.] [ABSTRACT FROM AUTHOR]