Radiofrequency ablation (RFA) is increasingly being used to treat unresectable liver tumors. Complete ablation of the tumor and a safety margin is necessary to prevent local recurrence. With current electrodes, size and shape of the ablation zone are highly variable leading to unsatisfactory local recurrence rates, especially for tumors >3 cm. In order to improve predictability, we recently developed a system with four simple electrodes with complete ablation in between the electrodes. This rather small but reliable ablation zone is considered as a building block for matrix radiofrequency ablation (MRFA). In the current study we explored the influence of the electric mode (monopolar or bipolar) and the activation mode (consecutive, simultaneous or switching) on the size and geometry of the ablation zone.