Study on Temperature Characteristics of Capacitance Diaphragm Gauge
- Resource Type
- Conference
- Authors
- Wu, Chengyao; Cheng, Yongjun; Sun, Wenjun; Pei, Xiaoqiang; Yang, Qihao; Ran, Xin
- Source
- 2023 24th International Vacuum Electronics Conference (IVEC) Vacuum Electronics Conference (IVEC), 2023 24th International. :1-2 Apr, 2023
- Subject
- Components, Circuits, Devices and Systems
Engineered Materials, Dielectrics and Plasmas
Temperature measurement
Temperature sensors
Heating systems
Measurement errors
Temperature distribution
Interpolation
Capacitance-voltage characteristics
Capacitance diaphragm gauge
Temperature characteristic
Interpolation of Delaunay
- Language
Capacitance Diaphragm Gauge (CDG) is a type of sensor commonly used under low vacuum with high accuracy and stability. Temperature is one of the most important factors affecting its measurement accuracy. When the ambient temperature changes, the measurement results of CDGs will also drift accordingly. We experimented with the effect of temperature on CDGs and quantified the measurement errors of heated and unheated CDGs in the environment of $\mathbf{(-30\sim50)} ^{\circ} \mathbf{C}$. The error caused by the temperature for the unheated CDG is relatively large. The temperature-pressure error surface we proposed in this paper can effectively reduce the measurement error of the unheated CDG and improve its accuracy.