Parametric Analysis for the effect of expansion ratio during Co-Cr stent manufacturing using fiber Laser.
- Resource Type
- Article
- Authors
- Patel, Jatin Haribhai; Patel, Dhaval M.
- Source
- Journal of Positive School Psychology; 2022, Vol. 6 Issue 3, p1307-1315, 11p
- Subject
- Computer software development
Manufacturing processes
Mathematical models
Coronary artery stenosis
Fiber lasers
- Language
- ISSN
- 27177564
This research shows how expansion ratio gets affected by the different input parameters of fiber laser. In this research all input parameters were studied and scrutinize the final parameters which affects or influence the mechanical properties of the stent. In this research the input parameters were selected after performing pilot experiments. ANOVA analysis is used to finding out the percentage contribution of each factors on expansion ratio. Mathematical model has also been derived for prediction of expansion ratio without performing actual experiments. It is observed less effect of gas pressur, contributed 10.36%, Percentage contribution of laser power is 61.88 % and frequency is contributing 26.64%. Software developed featured with which can predict expansion ratio of stent without performing actual performance. [ABSTRACT FROM AUTHOR]