This paper presents a wearable skin patch for wireless measurements of protein biomarkers in dermal interstitial fluid (ISF). ISF is extracted from the skin using a microneedle (MN)-based, vacuum-assisted technique and autonomously transported through the patch via vacuum pressure. This device was used for quantitative measurements of C-X-C motif chemokine ligand 9 (CXCL9), a biomarker for autoimmune diseases and inflammation, which could be detected from 10 pg/mL to 1,000 pg/mL in phosphate-buffered saline (PBS) with a lower limit of detection of 1.33 pg/mL. Proof-of-concept was demonstrated by performing measurements on cadaver porcine skin dermally injected with ISF simulant spiked with CXCL9, which could be detected at 100 and 1,000 pg/mL, thus validating the functionality of this wearable sensor.