Performance of 2 inch and 3 inch Scintillation Detectors with SiPM Light Readout
- Resource Type
- Conference
- Authors
- Szczesniak, T.; Grodzicka-Kobylka, M.; Moszynski, M.; Szawlowski, M.; Mianowski, S.; Wolski, D.
- Source
- 2017 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2017 IEEE. :1-4 Oct, 2017
- Subject
- Bioengineering
Components, Circuits, Devices and Systems
Computing and Processing
Nuclear Engineering
Photonics and Electrooptics
Detectors
Energy resolution
Scintillators
Crystals
Linearity
Resistance
Photomultipliers
MPPC
Scintillation detectors
SiPM
- Language
- ISSN
- 2577-0829
Spectroscopic performance of 4 types of large detectors consisting of 3 different scintillators (LaBr 3 , NaI(Tl) and BGO) coupled to large MPPC array are presented. The measurements are made with S12642–1616PB-50(X) Hamamatsu TSV MPPC array having a 50x50 µm 2 cell size and effective active area of 48x48 mm 2 (16x16 channels). In all measurements the common readout of all channels is used. The results obtained with MPPC, such as energy resolution and linearity, are compared with measurements of the same crystals on classic photomultiplier. Also the possible optimization of data acquisition electronics is presented.