A lack of comparative data across laboratories is often a barrier to the uptake and adoption of new technologies. Furthermore, data generated by different immunoassay methods may be incomparable due to a lack of harmonization. In this multicenter study, we describe validation experiments conducted in a single lab and cross-lab comparisons of assay results to assess the performance characteristics of the Q-plex™ 7-plex Human Micronutrient Array (7-plex), an immunoassay that simultaneously quantifies seven biomarkers associated with micronutrient (MN) deficiencies, inflammation and malarial antigenemia using plasma or serum; alpha-1-acid glycoprotein, C-reactive protein, ferritin, histidine-rich protein 2, retinol binding protein 4, soluble transferrin receptor, and thyroglobulin. Validations included repeated testing (n = 20 separately prepared experiments on 10 assay plates) in a single lab to assess precision and linearity. Seven independent laboratories tested 76 identical heparin plasma samples collected from a cohort of pregnant women in Niger using the same 7-plex assay to assess differences in results across laboratories. In the analytical validation experiments, intra- and inter-assay coefficients of variation were acceptable at