Validation of a novel particle isolation procedure using particle doped tissue samples
- Resource Type
- article
- Authors
- J. Patel; S. Lal; S.P. Wilshaw; R.M. Hall; J.L. Tipper
- Source
- Data in Brief, Vol 18, Iss , Pp 1802-1807 (2018)
- Subject
- Computer applications to medicine. Medical informatics
R858-859.7
Science (General)
Q1-390
- Language
- English
- ISSN
- 2352-3409
A novel particle isolation method for tissue samples was developed and tested using particle-doped peri-articular tissues from ovine cadavers. This enabled sensitivity of the isolation technique to be established by doping tissue samples of 0.25 g with very low particle volumes of 2.5 µm3 per sample. Image analysis was used to verify that the method caused no changes to particle size or morphologies.