In this work a hydrophilic interaction liquid chromatography/positive ion electrospray mass spectrometric assay (HILIC/ESI-MS) has been developed and fully validated for the quantitation of metformin and rosuvastatin in human plasma. Sample preparation involved the use of 100 &mu
L of human plasma, following protein precipitation and filtration. Metformin, rosuvastatin and 4-[2-(propylamino) ethyl] indoline 2 one hydrochloride (internal standard) were separated by using an X-Bridge-HILIC BEH analytical column (150.0 ×
2.1 mm i.d., particle size 3.5 &mu
m) with isocratic elution. A mobile phase consisting of 12% (v/v) 15 mM ammonium formate water solution in acetonitrile was used for the separation and pumped at a flow rate of 0.25 mL min&minus
1. The linear range of the assay was 100 to 5000 ng mL&minus
1 and 2 to 100 ng mL&minus
1 for metformin and rosuvastatin, respectively. The current HILIC-ESI/MS method allows for the accurate and precise quantitation of metformin and rosuvastatin in human plasma with a simple sample preparation and a short a chromatographic run time (less than 15 min). Plasma samples from eight patients were further analysed proving the capability of the proposed method to support a wide range of clinical studies.